Technical Specification

Acuatic 12 B - overall dimensions
Specification Attributes Values
Dimensions Weight
Chamber Diameter
Chamber Length
Chamber Volume
23kg
220mm
320mm
12 litres
Trays Number
Width
Length
Maximum loading per tray
4
184mm
284mm
1.5kg
Clearances Top and sides
Rear
50mm
None required
Cycle Type B-type vacuum
N-type non vacuum
For any type of instrument - wrapped or unwrapped
For solid unwrapped instruments
Cycle Time N-type
B-type
12 to 20 minutes*
20 to 30 minutes*
Sterilising Temperature
Time
134°C to 138°C
3 minutes minimum
Limits Design pressure
Design temperature
-1 to 3 bar
2 to 144°C
Display Resolution Temperature
Pressure
0.1°C
0.01 bar
Display Accuracy Between 75 to 150°C
Between 134 and 138°C
± 2°C
+ 0/ -1.5°C
Electrical Voltage
Current
Power
Frequency
Plug top fuse
Internal fuse
Energy used per cycle
230 V ac ± 10%
Variable up to 13A
Variable up to 3 kW
50Hz
13 A
1 A
0.2 to 0.7 kWh*
Water Water used per cycle
Clean water reservoir volume
Used water reservoir volume
200 to 700 ml*
2 litre
2 litre
Cycle Recording Type
Recording media
Capacity
Inbuilt 'Securilog' data logger
SD Memory card
Life time of steriliser
Certification Council Directive 93/42/EEC Medical Devices

* Cycle time and electrical and water usage and dependent on steriliser loading and cycle start temperature.